Specifications:
Chamber Volume2.75 cu. ft.
Temperature RangeAmbient +5 to 95°C
Temperature Uniformity 
@37°C±0.2°C
@56°C±0.5°C
@90°C±0.75°C
Temperature Accuracy±0.2°C
Temperature Display "D"to 0.1°C
Temperature Display "R"to 1.0°C
Dimensions (WxDxH inches) :
Chamber17x14x20
Outer27x20x29
Electrical120V-60Hz (800W)
Recovery Time
(measured with fan on after door opened for 30 seconds)
 
@37°C55 seconds
@56°C6 minutes

These High Performance Digitally controlled incubators are used for applications requiring a wide range of temperatures from ambient +5°C to 95°C, allowing general incubation (<65°C) and experiments requiring higher temperature. Feature a digital microprocessor temperature control to assure accurate and reliable temperature stability. This control is designed to be set and displayed digitally for "set and forget" accuracy and peace of mind.

True digital / digital temperature control

The temperature is selected and displayed digitally. Enter a temperature by pressing the "up" or "down" arrow, and walk away.

Switchable turbo fan

A high performance fan assists the radiant heat distribution system. Fro rapid temperature changeover, simply press the fan switch to activate the turbo fan option. The fan increases circulation permitting rapid temperature increase and decrease. When the fan is activated, the forced circulation provides uniformity that is superior to standard mode radiant heat distribution.

Fail-Safe over temperature protection

An independent hydraulic safety thermostat protects samples and equipment from excessive heat in the event of a primary control failure. All electronics are located away from the heated chamber, protected from heat related stresses by a cooling fan (within the electronics module) which continuously bathes all components in ambient air, ensuring longer life of electronic control components.

Interior electrical outlet and illumination

Inside the stainless steel chamber, you'll find a recessed light source, allowing visual examination of samples through the gasketed inner glass door. The light is on only when the outer door is opened. An electrical outlet is provided for operation of optional equipment (rockers, shakers, etc.) inside the heated chamber.

Heated outer door

Warm air is evenly distributed from the bottom, the back, and both sides of the chamber. The outer door is also heated, enclosing the entire chamber in a warm air "jacket", eliminating localized hot spots. The heated door also prevents condensation on the glass door. The outer door's magnetic gasket automatically forms an airtight seal.

Adjustable shelves and leveling feet

Supplied with 2 adjustable stainless steel shelves and leveling feet to compensate for uneven bench surfaces.

311D Digital Incubator

Chamber Volume: 2.75 cu. Ft.
Temperature Range: Ambient +5 to 95°C
SKU: LI-5311-D
Old price: $3,885.30