Calibration Check™ pH/mV/Temperature Bench Meter
- One or two input channels
- pH Calibration Check™
- Five point pH calibration with seven standard and five custom buffers
- Stability, interval and log on demand logging
- Up to 400 log on demand records and 600 automatic logging records
- Messages on the graphic LCD for an easy and accurate calibration
- Multi-language support
- GLP features
- PC interface via USB
HANNA’s HI 3220 benchtop instrument features up to five point pH calibration with a choice of five custom buffers and seven standard buffers.
HI 3220 features an interactive user support interface that assists you before, during and after measurement. On-screen tutorials guide users through set-up, calibration and measurement while context sensitive help of any screen is available at a push of a button. The help screen includes language specific assistance for menu parameters, calibration, logging, contact and accessory information for your instrument.
This instrument features HANNA’s exclusive Calibration Check™, a diagnostics system that ensures accurate pH readings every time. By alerting users of potential problems during the calibration process, the Calibration Check™ system eliminates erroneous readings due to dirty or faulty pH electrodes or contaminated pH buffer solutions during calibration. Throughout the calibration process, users are guided step-by-step by the on-screen tutorial. After calibration, the probe condition is evaluated and an indicator is displayed informing the user of the overall pH electrode status.
The HI 3220 is equipped with one input channel.
This instrument can measure using ORP electrodes (pH channel input), thanks to it's capability to measure mV with a resolution up to 0.1 mV.
Order Information:
HI 3220-01 (115V) and HI 3220-02 (230V), is supplied with HI 1131B pH electrode*, HI 7662-T temperature probe, HI 76404N electrode holder, HI 70004 pH 4.01 buffer solution sachet, HI 70007 pH 7.01 buffer solution sachet, HI 700661 electrode cleaning solution sachet (2), HI 7071S electrolyte solution (30 mL), 12 VDC adapter and instructions.
*TRIS compatible meter includes HI1043B electrode.
*HI3220TRIS includes TRIS compatible electrode (HI1043B) in place of HI1131B
Range | pH | -2.0 to 20.0; -2.00 to 20.00; -2.000 to 20.000 pH |
| ORP | ±2000 mV |
| Temperature | -20.0 to 120.0 °C (-4.0 to 248.0 °F) |
Resolution | pH | 0.1; 0.01; 0.001 pH |
| ORP | 0.1 mV |
| Temperature | 0.1 °C (0.1 °F) |
Accuracy | pH | ±0.01; ±0.002 pH |
| ORP | ±0.2 mV |
| Temperature | ±0.2°C (±0.4°F) (excluding probe error) |
Calibration | pH | Up to 5 point calibration, 7 standard buffers available (1.68, 4.01, 6.86, 7.01, 9.18, 10.01, 12.45) + 5 custom buffers |
Calibration | Slope | From 80 to 110% |
Temperature Compensation (pH) | | Manual or automatic from -20.0 to 120.0 °C(-4.0 to 248.0 °F) |
Probes | | HI 1131B pH with BNC connector and HI 7662-T stainless steel temperature probe (included) |
Logging | | Log-on-demand 200 samples |
PC Connectivity | | Opti-isolated USB (with HI 92000 software) |
Lot Logging | | 5, 10, 30 sec 1, 2, 5, 10, 15, 30, 60, 120, 180 min |
Input Impedance | | 1012 |
Power Supply | | 12 VDC |
Environment | | 0-50 °C (32 to 122 °F) Max. RH 55% non-condensing |
Dimensions | | 235 x 207 x 110 mm (9.2 x 8.14 x 4.33") |
Weight | | 1.8 kg (4.1 lbs.) |